Calibration services

The Metrology Research Institute is the Finnish National Standards Laboratory for optical quantities. MRI gives calibrations on the optical quantities listed in the table below with the indicated measurement ranges and uncertainties.

The Metrology Research Institute also offers consultation services and other support in measurement technology.

Contact: farshid.manoocheri [at] aalto [dot] fi

 

QUANTITIES, RANGES, WAVELENGTH RANGES AND UNCERTAINTIES

Quantity

Range

Wavelength range

 (k=2)
Uncertainty

Luminous intensity

10 – 10 000 cd

0,3 %

 

10 – 200 lx

0,2 %

Illuminance

200 – 2 000 lx

 

0,3 % – 0,5 %

Luminance

5 – 40 000 cd m-2

0,8 %

Luminous flux

10 – 10 000 lm

1,0 %

 

10 pW mm-2 nm-1 –

290 – 380 nm

1,1 % – 2,8 %

Spectral irradiance

500 mW mm-2 nm-1

380 – 900 nm

0,6 % – 1,1 %


Spectral radiance

5 mW m-2 sr-1 nm-1 – 6 W m-2 sr-1 nm-1

290 – 900 nm

1,0 %

Colour coordinates
(x, y)

0 – 1

0,1 %

Colour temperature

1 000 – 3 500 K

0,15 %

Optical power

0,1 – 0,5 mW

325, 442, 457.9, 465.8, 472.7, 476.5, 482.6, 488.1, 496.7, 501.7, 514.5, 520.9, 531, 543.5, 568.4, 633, 635, 647.3, 676.6, 932, 1523 nm

0,05 %

 

10 nW – 1 mW

250 – 380 nm

0,3 % – 0,8 %

 

10 nW – 1 mW

380 – 920 nm

0,1 % – 0,3 %

 

10 mW - 10 W

250 nm – 16 mm

2 % – 10 %


QUANTITIES, RANGES, WAVELENGTH RANGES AND UNCERTAINTIES

Transmittance

0,0005 – 1

250 – 380 nm

0,1 % – 1 %

 

0,0001 – 1

380 – 920 nm

0,05 % – 0,5 %

 

0,0005 – 1

920 – 1700 nm

0,1 % – 1 %

Spectral responsivity

 

250 – 380 nm

1 %

 

> 0,01 A/W

380 – 920 nm

0,5 %

 

 

920 – 1700 nm

4 %

Reflectance4 (5°–85°)

0,05 – 1

250 – 380 nm

0,5 % – 5 %

 

0,01 – 1

380 – 1000 nm

0,3 % – 3 %

Diffuse reflectance factor4

0,05 – 1

360 – 830 nm

0,4 % – 1 %

Optical wavelength

400 nm – 1,55 mm

0,01 nm

 

 

 

 

Page content by: | Last updated: 20.08.2014.